Numerical Simulation of Nonpulsatile Left Ventricular Bypass
- Source :
- Artificial Organs. 20:632-636
- Publication Year :
- 1996
- Publisher :
- Wiley, 1996.
-
Abstract
- A computer simulation was carried out to investigate the influence of nonpulsatile left ventricular assistance on hemodynamics. A simulation circuit was constructed to represent the circulatory system. A source of current was added to denote the nonpulsatile blood pump. The left and right ventricles were replaced by variable compliances. Left heart failure was simulated by decreasing the amount of compliance change of the left ventricle. We introduced a pulsatility indicator (PI) to clarify the pulsatility characteristics in the hemodynamics; this PI was defined as the ratio of the pulse pressure (PP) to the mean aortic pressure (AoP). When nonpulsatile bypass flow increased, the mean AoP, tension time index (TTI), and diastolic pressure time index (DPTI) increased, and cardiac output, PP, and PI decreased. When assisted flow increased with the constant total flow rate, the mean AoP and DPTI changed little; the PP, TTI, and PI decreased, and the endocardial viability rate increased. The PI would be helpful in evaluating the effect of pulsatility.
- Subjects :
- Heart Bypass, Left
medicine.medical_specialty
Cardiac output
Biomedical Engineering
Medicine (miscellaneous)
Hemodynamics
Blood Pressure
Bioengineering
Biomaterials
Internal medicine
Medicine
Computer Simulation
Cardiac Output
Heart Failure
business.industry
Mean Aortic Pressure
General Medicine
Models, Theoretical
medicine.disease
Pulse pressure
Blood pump
medicine.anatomical_structure
Blood pressure
Ventricle
Pulsatile Flow
Heart failure
Cardiology
Rheology