Velocity encoding and velocity compensation for multi-spoke RF excitation

Authors :
Peter Bachert
Mark E. Ladd
Sebastian Schmitter
Sebastian Flassbeck
Simon Schmidt
Source :
Magnetic Resonance Imaging. 66:69-85
Publication Year :
2020
Publisher :
Elsevier BV, 2020.

Abstract

Purpose To investigate velocity encoded and velocity compensated variants of multi-spoke RF pulses that can be used for flip-angle homogenization at ultra-high fields (UHF). Attention is paid to the velocity encoding for each individual spoke pulse and to displacement artifacts that arise in Fourier transform imaging in the presence of flow. Theory and methods A gradient waveform design for multi-spoke excitation providing an algorithm for minimal TE was proposed that allows two different encodings. Such schemes were compared to an encoding approach that applies an established scheme to multi-spoke excitations. The impact on image quality and quantitative velocity maps was evaluated in phantoms using single- and two-spoke excitations. Additional validation measurements were obtained in-vivo at 7 T. Results Phantom experiments showed that keeping the first gradient moment constant for all k-space lines eliminates any displacements in phase-encoding and slice-selection direction for all spoke pulses but leads to artifacts for non-zero velocity components along readout direction. Introducing variable but well-defined first gradient moments in the phase-encoding direction creates displacements along the velocity vector and thus minimizes velocity-induced geometrical distortions. Phase-resolved mean volume flow in the ascending and descending aorta obtained from two-spoke excitation showed excellent agreement with single-spoke excitation over the cardiac cycle (mean difference 0.8 ± 16.2 ml/s). Conclusions The use of single- and multi-spoke RF pulses for flow quantification at 7 T with controlled displacement artifacts has been successfully demonstrated. The presented techniques form the basis for correct velocity quantification and compensation not only for conventional but also for multi-spoke RF pulses allowing in-plane B1+ homogenization using parallel transmission at UHF.

Details

ISSN :
0730725X
Volume :
66
Database :
OpenAIRE
Journal :
Magnetic Resonance Imaging
Accession number :
edsair.doi.dedup.....7bd98b2d97c447b886e9a30f9d346b71
Full Text :
https://doi.org/10.1016/j.mri.2019.11.007