Back to Search Start Over

Atrial arrhythmia suppression by atrial overdrive pacing: pacemaker Holter assessment

Authors :
Cliff Bucknall
K J Ward
Jaswinder Gill
Kayvan Kamalvand
J E Willett
Source :
Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology. 3(2)
Publication Year :
2001

Abstract

The purpose of this study was to assess the efficacy of overdrive, single-site right atrial appendage pacing to reduce the burden of atrial fibrillation (AF) when compared with a standard lower rate limit of 60 bpm. This was verified by using the pacemaker's Holter. Eighteen subjects with a pre-implant history of paroxysmal AF and implanted DDDR mode-switching pacemakers were recruited. The pacemaker lower rate limit was programmed in random order to 60, 75 or 90 bpm for three 2-month periods and the amount of AF quantified. In addition, the exercise tolerance (ET), general well being (GWB), functional capacity (FC) and specific symptom prevalence (SSP) were assessed. The main finding of the study was that when ranked according to the amount of AF, there was no significant difference in the amount of AF according to the pacing rate. Six patients failed to tolerate pacing at 90 bpm. There were no differences in the ET, GWB, FC and SSP scores. It is concluded that those clinicians that manage patients with paroxysmal

Details

ISSN :
10995129
Volume :
3
Issue :
2
Database :
OpenAIRE
Journal :
Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology
Accession number :
edsair.doi.dedup.....7e8fe8c5f9fb249943102b37bd9dd968