Comparison of nonuniform rotational distortion between mechanical IVUS and OCT using a phantom model
- Source :
- Ultrasound in Medicine & Biology. 33:67-73
- Publication Year :
- 2007
- Publisher :
- Elsevier BV, 2007.
-
Abstract
- Optical coherence tomography (OCT) is an optical analog of mechanical intravascular ultrasound (M-IVUS) with much higher spatial resolution. However, no data exist regarding the nonuniform rotational distortion (NURD) with OCT. The aim of the study was to investigate whether OCT generates less NURD relative to M-IVUS. A coronary artery phantom model was constructed with a rubber ring (3.68 mm in diameter), located at the distal end of the phantom. This model was also composed of eight equally spaced steel wires and an additional marker-wire. Two types of vascular phantoms were used, mild curve (90 degrees ) and acute curve (near 180 degrees ). Subsequent M-IVUS (n = 6) and OCT (n = 6) imaging was performed. Eight angles between eight wires, except the marker-wire, were measured from each image. These angles, measured with M-IVUS and OCT, were compared with those of high-resolution optical photography as a gold standard. The average in angle differences was significantly smaller in OCT compared with M-IVUS in the mild curve model (3.2 +/- 1.0 degrees vs, 6.9 +/- 2.1 degrees , p < 0.01). Compared with the latter model, the average in angle differences was exaggerated in the acute curve model with M-IVUS (9.1 +/- 0.9 degrees vs. 6.9 +/- 2.1 degrees , p < 0.05) but not with OCT (3.5 +/- 0.8 degrees vs. 3.2 +/- 1.0 degrees , p= not significant). OCT generates significantly less NURD compared with M-IVUS, especially in tortuous situation.
- Subjects :
- Materials science
Rotation
genetic structures
Acoustics and Ultrasonics
Biophysics
Imaging phantom
Catheterization
Optics
Nuclear magnetic resonance
Optical coherence tomography
Distortion
Intravascular ultrasound
medicine
Humans
Radiology, Nuclear Medicine and imaging
Image resolution
Ultrasonography, Interventional
Rubber ring
Radiological and Ultrasound Technology
medicine.diagnostic_test
Phantoms, Imaging
business.industry
Equipment Design
equipment and supplies
Coronary Vessels
eye diseases
sense organs
Tomography
Artifacts
business
Tomography, Optical Coherence
Subjects
Details
- ISSN :
- 03015629
- Volume :
- 33
- Database :
- OpenAIRE
- Journal :
- Ultrasound in Medicine & Biology
- Accession number :
- edsair.doi.dedup.....823e7ffccba7169d1f9eb2496ae1ca94
- Full Text :
- https://doi.org/10.1016/j.ultrasmedbio.2006.07.020