Experimental determination of the effective point of measurement for cylindrical ionization chambers in megavoltage photon beams

Authors :
Yukiko Iwafuchi
Hiroshi Oguchi
Kuniyasu Okudaira
Kota Yamamoto
Source :
Radiological Physics and Technology. 15:291-297
Publication Year :
2022
Publisher :
Springer Science and Business Media LLC, 2022.

Abstract

Current dosimetry protocols specify an effective point of measurement (EPOM) shift of 0.6r for a cylindrical ionization chamber in photon beams. However, prior studies have reported that this shift was excessively large. The objective of this study was to experimentally evaluate the EPOM shifts in photon beams for cylindrical ionization chambers, which are widely used in clinical practice, and thus determine the appropriate EPOM shift. A microdiamond detector, which is a semiconductor detector with a small sensitive volume, was used as a reference detector, and the EPOM shifts of 11 types of cylindrical ionization chambers were evaluated at 6 MV and 10 MV. The depth shift from the percent depth dose (PDD) of the reference detector to that of the evaluated chamber was calculated using the least-squares method and was defined as the EPOM shift. The EPOM shift of the 10 MV condition was slightly larger than that of the 6 MV condition. However, because this trend was not observed for all chambers, the results of the two energies were averaged, and the EPOM shifts were determined to be 0.33r-0.43r (± 0.05) for 10 types of ionization chambers, and 0.03r (± 0.03) for the A1SL chamber. The shifts for all ionization chambers were smaller than 0.6r, indicating that the recommended EPOM shifts were overestimated and the absorbed dose was underestimated at the calibration depth. Hence, the appropriate EPOM shift of the 10 types of ionization chambers was 0.4r (the geometric center of the A1SL chamber), with a dose uncertainty of 0.05%.

Details

ISSN :
18650341 and 18650333
Volume :
15
Database :
OpenAIRE
Journal :
Radiological Physics and Technology
Accession number :
edsair.doi.dedup.....83e671779aaf231299d45e8eecd767e8
Full Text :
https://doi.org/10.1007/s12194-022-00669-z