Back to Search Start Over

Comparison of Fiber-Based Frequency Comb and Ti:Sapphire-Based Frequency Comb

Authors :
Sang Eon Park
Dae-Su Yee
Ho Suhng Suh
Won-Kyu Lee
Eok Bong Kim
Dai-Hyuk Yu
Chang Yong Park
Source :
2007 Conference on Lasers and Electro-Optics - Pacific Rim.
Publication Year :
2007
Publisher :
IEEE, 2007.

Abstract

For the first time we compare two kinds of optical frequency combs, one of which is based on a Ti:sapphire femtosecond laser and the other is based on a mode-locked erbium-doped fiber laser. The comparison is performed by measuring an optical frequency standard with these two combs simultaneously. The two frequency measurements agree within 1.8 Hz ( $3.8{\times}10^{-15}$ ) with the uncertainty of 17.2 Hz ( $3.6{\times}10^{-14}$ ), from which it can be concluded that the Ti:sapphire-based frequency comb and the fiber-based frequency comb have no systematic discrepancy at this level of uncertainty.

Details

Database :
OpenAIRE
Journal :
2007 Conference on Lasers and Electro-Optics - Pacific Rim
Accession number :
edsair.doi.dedup.....84f30c124f27cff05097bf06cdf690de
Full Text :
https://doi.org/10.1109/cleopr.2007.4391797