Chemosaturation Percutaneous Hepatic Perfusion (CS-PHP) with Melphalan: Evaluation of 2D-Perfusion Angiography (2D-PA) for Leakage Detection of the Venous Double-Balloon Catheter

Authors :
Frank Wacker
Timo C Meine
Roman Kloeckner
Sabine K Maschke
Cornelia L A Dewald
Hinrich B. Winther
Julius Renne
Jan B. Hinrichs
Arndt Vogel
Bernhard C. Meyer
Martha M. Kirstein
Source :
Cardiovascular and interventional radiology. 42(10)
Publication Year :
2019

Abstract

To evaluate the feasibility of 2D-perfusion angiography (2D-PA) for detecting leakage of the double-balloon catheter used for chemosaturation percutaneous hepatic perfusion (CS-PHP). Overall, 112 CS-PHP (09/2015-09/2018) in 52 patients were retrospectively screened for leakage alongside the double-balloon catheter on standard venograms. Finally, 18 procedures with visually detected leakage were included. Fifteen consecutive procedures without leakage served as control. To evaluate 2D-PA for leakage detection, the acquired digital subtraction venograms were post-processed. For each balloon, two different target ROIs were evaluated to assess a possible impact of localization and shape of the ROIs. Time to peak (TTP), peak density (PD), area under the curve (AUC), and ratios of target ROI/reference ROIs (PDtROI/PDREF; AUCtROI/AUCREF; and TTPtROI/TTPREF) were calculated. Leakages were located as follows: 15/18 cranial and 3/18 caudal. At the cranial balloon both ROIs showed a significant decrease in PDtROI/PDREF and AUCtROI/AUCREF (ROI1: p

Details

ISSN :
1432086X
Volume :
42
Issue :
10
Database :
OpenAIRE
Journal :
Cardiovascular and interventional radiology
Accession number :
edsair.doi.dedup.....865b2d260248282f62c84fbb74087c94