Abstract

Background The extent of aortic valve calcification is an important determinant of procedural success in transcatheter aortic valve implantation (TAVI). We sought to validate device landing zone calcium volume (DLZ-CV) measurements on contrast-enhanced multidetector computed tomography (MDCT) with non-contrast-enhanced scans as reference. Methods We determined DLZ-CV in 141 patients undergoing transfemoral TAVI. Non-contrast-enhanced images were analyzed using a threshold of 130 HU as reference (DLZ-CV130). For contrast-enhanced scans, we applied various thresholds including 450 HU (DLZ-CV450), 850 HU (DLZ-CV850), mean aortic attenuation (AttenAo) + 2 SD (DLZ-CV2SD), AttenAo + 4 SD (DLZ-CV4SD), AttenAo + 4 SD + 5 mm3 volume filter (DLZ-CV4SD+), and based on visual estimation (DLZ-CVvis). We compared DLZ-CV values between patients with versus without paravalvular leak (PVL), and between patients with versus without post-dilatation stratified by the type of prosthesis. Results All DLZ-CV measurements on contrast-enhanced scans significantly differed from DLZ-CV130 (p Conclusion Measurement of DLZ-CV using contrast-enhanced scans with unadjusted thresholds results in incorrect estimation of the calcium volume. The use of a scan-specific individual HU threshold including a volume filter (DLZ-CV4SD+) provides the best approximation to the reference and allows for discrimination of PVL ≥ 1° in patients receiving the Acurate neo prosthesis.

Details

ISSN :
01675273
Volume :
263
Database :
OpenAIRE
Journal :
International Journal of Cardiology
Accession number :
edsair.doi.dedup.....8b0e00fe1d8f5c19c723e40d7f01bbf7
Full Text :
https://doi.org/10.1016/j.ijcard.2018.02.042