Continuous cardiac output monitoring with an uncalibrated pulse contour method in patients supported with mechanical pulsatile assist device
- Source :
- Interactive cardiovascular and thoracic surgery. 13(1)
- Publication Year :
- 2011
-
Abstract
- We evaluated the accuracy of an uncalibrated pulse contour method called Pressure Recording Analytical Method (PRAM) compared with continuous thermodilution for cardiac output (CO) monitoring in patients implanted with a pulsatile left ventricular assist device (LVAD). Twelve adult patients implanted with the HeartMate I-XVE device were studied. CO was simultaneously evaluated by PRAM and by continuous thermodilution. Blood flow values displayed by the LVAD's console were also used for the comparison. Bland-Altman and linear regression analyses were applied. A total of 72 CO measurements (range 3.8-6.7 l/min) were obtained. Mean CO was 5.23±0.70 l/min for the 'hot' pulmonary thermodilution (ThD-CCO) method, 5.28±0.63 l/min for PRAM and 4.83±0.67 l/min for LVAD-CO. A high correlation (r=0.90), a good agreement (mean bias -0.04 l/min, precision ±0.38 l/min) and a low percentage of error (7.3%) were observed between PRAM-CO and ThD-CCO. A good correlation was found between LVAD-CO and either ThD-CCO (r=0.88) or PRAM-CO (r=0.86), but an overestimation of 10% was observed for both PRAM-CO (mean bias -0.44 l/min) and ThD-CCO (mean bias -0.40 l/min). Our results demonstrated good agreements between PRAM-CO, ThD-CCO and LVAD-CO. PRAM derives CO from a peripheral artery without calibration and may be a complementary tool in the hemodynamic assessment of patients supported with a VAD.
- Subjects :
- Male
Cardiac output
Time Factors
medicine.medical_treatment
cardiac output
heart failure
pulse contour analysis
thermodilution technique
ventricular assist device
Thermodilution
Pulsatile flow
Hemodynamics
Blood Pressure
Ventricular Function, Left
Cardiac Output
Pulse (signal processing)
Signal Processing, Computer-Assisted
Pulse contour analysis
Middle Aged
Intensive Care Units
Treatment Outcome
Italy
Radial Artery
Ventricular assist device
Cardiology
Female
Cardiology and Cardiovascular Medicine
Pulmonary and Respiratory Medicine
medicine.medical_specialty
Pulmonary Artery
Predictive Value of Tests
Internal medicine
Linear regression
Catheterization, Peripheral
medicine
Calibration
Humans
Aged
Monitoring, Physiologic
Heart Failure
Thermodilution technique
business.industry
Reproducibility of Results
Blood Pressure Determination
Blood flow
Blood Pressure Monitors
Regional Blood Flow
Catheterization, Swan-Ganz
Linear Models
Surgery
Heart-Assist Devices
Heart failure, Ventricular assist device, Cardiac output, Pulse contour analysis, Thermodilution technique
business
Biomedical engineering
Subjects
Details
- ISSN :
- 15699285
- Volume :
- 13
- Issue :
- 1
- Database :
- OpenAIRE
- Journal :
- Interactive cardiovascular and thoracic surgery
- Accession number :
- edsair.doi.dedup.....8c4072babf0f67b6c860f3713e23c134