Back to Search Start Over

Effect of fluid pressure on Coriolis mass flowmeter's performance

Authors :
Furio Cascetta
Cascetta, Furio
Source :
Scopus-Elsevier
Publication Year :
1996
Publisher :
Elsevier BV, 1996.

Abstract

This paper presents some tests which have been carried out on a commercial model of Coriolis mass flowmeter under medium-pressure conditions ( p > 15 bar), utilizing a primary standard flow calibration facility. The experimental results show that the fluid pressure affects the accuracy of the tested meter, which provides underestimated readings (i.e. negative errors). Finally, some diagrams which allow to correct the output signal (or the reading) provided by the tested meter (according to the operating fluid pressure) are also presented.

Details

ISSN :
00190578
Volume :
35
Database :
OpenAIRE
Journal :
ISA Transactions
Accession number :
edsair.doi.dedup.....8d7314d82644a8532663a6ec34d66d84
Full Text :
https://doi.org/10.1016/s0019-0578(96)00048-1