P.47 Feasibility Evaluation of Imaging-Free Ultrasound Technology to Measure Diameters of Brachial and Radial Arteries for Assessment of Endothelial Function
- Source :
- Artery Research, Vol 26, Iss Supplement 1 (2020)
- Publication Year :
- 2020
- Publisher :
- Atlantis Press, 2020.
-
Abstract
- Background: Ultrasonographic imaging to record changes in peripheral arterial diameter associated with Flow mediated dilatation or Low flow mediated constriction is routinely used to assess various facets of vascular endothelial function. Imaging poses many challenges including requirement of costly ultrasound machines, trained manpower to perform imaging and effort-intensive steps to analyse the images subsequently using manual or automated methods. We tested the feasibility and validity of using an imaging-free technology to record resting arterial diameters of brachial and radial arteries. Methods: Eight healthy volunteers initially underwent ultrasonographic imaging (M7, Mindray; Shenzhen, P.R. China) of brachial artery and proximal radial artery. The brachial and radial artery ‘zones’ thereby identified through imaging were surface marked on subject’s arm. Imaging-free ARTSENS® Pen device [1] (Healthcare Technology Innovation Centre, IIT Madras, India) consisting of highly integrated hardware for operating a single element broadband ultrasound transducer (centre-frequency = 5 MHz, spatial half angle
Details
- Language :
- English
- ISSN :
- 18764401
- Volume :
- 26
- Database :
- OpenAIRE
- Journal :
- Artery Research
- Accession number :