- We have demonstrated a novel Fourier-domain optical coherence tomography system and signal-processing algorithm for full-range, real-time, artifact-free quantitative imaging of the anterior chamber. Cross-sectional full-range images comprising 1024 x 800 pixels (axial x lateral) were acquired and displayed at 6.7 images/s. Volumetric data comprising 1024 x 400 x 60 pixels (axial x lateral x elevation) were acquired in 4.5 seconds with real-time visualization of individual slices and 3-dimensional reconstruction performed in postprocessing. Details of the cornea, limbus, iris, anterior lens capsule, trabecular meshwork, and Schlemm's canal were visualized. Quantitative surface height maps of the corneal epithelium and endothelium were obtained from the volumetric data and used to generate corneal thickness maps.
- Subjects :
- Materials science
genetic structures
Pilot Projects
Article
Optics
Optical coherence tomography
Anterior Eye Segment
Computer Systems
Cornea
Image Processing, Computer-Assisted
medicine
Medical imaging
Humans
Iris (anatomy)
Corneal epithelium
Anatomy, Cross-Sectional
Fourier Analysis
medicine.diagnostic_test
business.industry
Corneal Topography
Corneal topography
eye diseases
Ophthalmology
medicine.anatomical_structure
sense organs
Trabecular meshwork
business
Algorithms
Tomography, Optical Coherence
Subjects
Details
- ISSN :
- 00039950
- Volume :
- 126
- Database :
- OpenAIRE
- Journal :
- Archives of Ophthalmology
- Accession number :
- edsair.doi.dedup.....93e4d6d43f6c1027a5691f28c4b5b5c7
- Full Text :
- https://doi.org/10.1001/archopht.126.4.537