Back to Search Start Over

Imaging the Ocular Anterior Segment With Real-Time, Full-Range Fourier-Domain Optical Coherence Tomography

Authors :
Joseph A. Izatt
Sanjay Asrani
Marinko V. Sarunic
Source :
Archives of Ophthalmology. 126:537
Publication Year :
2008
Publisher :
American Medical Association (AMA), 2008.

Abstract

We have demonstrated a novel Fourier-domain optical coherence tomography system and signal-processing algorithm for full-range, real-time, artifact-free quantitative imaging of the anterior chamber. Cross-sectional full-range images comprising 1024 x 800 pixels (axial x lateral) were acquired and displayed at 6.7 images/s. Volumetric data comprising 1024 x 400 x 60 pixels (axial x lateral x elevation) were acquired in 4.5 seconds with real-time visualization of individual slices and 3-dimensional reconstruction performed in postprocessing. Details of the cornea, limbus, iris, anterior lens capsule, trabecular meshwork, and Schlemm's canal were visualized. Quantitative surface height maps of the corneal epithelium and endothelium were obtained from the volumetric data and used to generate corneal thickness maps.

Details

ISSN :
00039950
Volume :
126
Database :
OpenAIRE
Journal :
Archives of Ophthalmology
Accession number :
edsair.doi.dedup.....93e4d6d43f6c1027a5691f28c4b5b5c7
Full Text :
https://doi.org/10.1001/archopht.126.4.537