Abstract
- Fast gas chromatography/negative-ion chemical ionization mass spectrometric (GC/NICI-MS) assay combined with rapid and nonlaborious sample preparation is presented for the simultaneous determination of benzodiazepines and alpha-hydroxy metabolites, zaleplon and zopiclone in whole blood. The compounds were extracted from 100 microl of whole blood by simultaneous multitube, microscale liquid-liquid extraction (LLE) and derivatized by N-methyl-N-(tert-butyldimethylsilyl)trifluoroacetamide (MTBSTFA), without the need for the time-consuming concentration stage. In the analytical separation, various parameters of fast GC/NICI-MS were applied, e.g. the use of hydrogen as a GC carrier gas, a high carrier gas velocity, a small film thickness of the analytical column, fast MS data acquisition, fast temperature ramping, and high initial and final temperatures of GC column. Sensitive identification, screening and quantitation of 18 compounds of interest were achieved in chromatographic separation in only 4.40 min. Accurate and reproducible results were obtained by using five different and carefully selected deuterated analogues on the basis of the chemical properties of the target analytes. Nevertheless, for alpha-OH-midazolam, and for bromazepam and flunitrazepam at low concentrations, the results can be considered only semiquantitative on the basis of the validation data. The extraction efficiencies ranged from 74.3 to 105.7% and the limits of quantitation (LOQ) from 1 to 100 ng ml(-1). Rapid sample preparation and fast chromatographic separation allowed cost-efficient, reliable and high sample-throughput analyses with a low amount of manual work. The method was fully validated and accredited according to EN ISO/IEC 17025 standards and is applicable for sensitive, reliable and quantitative determination of benzodiazepines, zaleplon and zopiclone, e.g. in clinical and forensic toxicology.
- Subjects :
- Spectrometry, Mass, Electrospray Ionization
Analyte
Time Factors
Mass spectrometry
Sensitivity and Specificity
Gas Chromatography-Mass Spectrometry
Piperazines
Specimen Handling
Benzodiazepines
chemistry.chemical_compound
Acetamides
Humans
Sample preparation
Derivatization
Spectroscopy
Chemical ionization
Chromatography
Chemistry
Microchemistry
Reproducibility of Results
Substance Abuse Detection
Pyrimidines
Gas chromatography
Gas chromatography–mass spectrometry
Azabicyclo Compounds
Quantitative analysis (chemistry)
Blood Chemical Analysis
Subjects
Details
- ISSN :
- 10969888 and 10765174
- Volume :
- 41
- Database :
- OpenAIRE
- Journal :
- Journal of Mass Spectrometry
- Accession number :
- edsair.doi.dedup.....96fcdd364caff6beea5147201ffb1f37
- Full Text :
- https://doi.org/10.1002/jms.1030