Differences between the wear couples metal-on-polyethylene and ceramic-on-ceramic in the stability against dislocation of total hip replacement
- Source :
- Journal of materials science. Materials in medicine. 15(6)
- Publication Year :
- 2004
-
Abstract
- After total hip replacement an insufficient range of motion (ROM) can lead to contact between femoral neck and rim of the cup (= impingement) causing dislocation and consecutive material failure. The purpose of this study was to analyse the influence of different wear couples on the ROM and stability against dislocation. By means of a special testing device the ROM until impingement, the ROM until dislocation as well as the resisting moment against levering the head out of the cup were experimentally determined. Various total hip systems with cup inserts made of ceramic and polyethylene were comparatively examined in different implant positions. Maximum resisting moment as well as the ROM until impingement and dislocation were clearly influenced by the implant position. Furthermore, the stability against dislocation was affected by design parameters, whereas in the case of appropriate implant position differing wear couples (metal-on-polyethylene vs. ceramic-on-ceramic) had a minor impact. However, as shown by tests under lubricant conditions, ceramic-on-ceramic couples provided less dislocation stability in unfavourable implant position in comparison to metal-on-polyethylene. Therefore, ceramic-on-ceramic couples should only be applied in the case of optimised implant orientation preventing impingement and dislocation with subsequent material failure like chipping off or breakage.
- Subjects :
- Joint Instability
Ceramics
Materials science
Friction
Biomedical Engineering
Biophysics
Bioengineering
Biomaterials
Breakage
Materials Testing
Hip Dislocation
Humans
Material failure theory
Ceramic
Lubricant
Composite material
Range of Motion, Articular
Metallurgy
Prosthesis Failure
Equipment Failure Analysis
Torque
Metals
Polyethylene
visual_art
visual_art.visual_art_medium
Head (vessel)
Hip Joint
Implant
Hip Prosthesis
Dislocation
Range of motion
Subjects
Details
- ISSN :
- 09574530
- Volume :
- 15
- Issue :
- 6
- Database :
- OpenAIRE
- Journal :
- Journal of materials science. Materials in medicine
- Accession number :
- edsair.doi.dedup.....9b0efdd420ad9ffc28a93ddc5ce234f9