Back to Search Start Over

Mechanical validation of an MRI compatible stereotactic neurosurgery robot in preparation for pre-clinical trials

Authors :
Julie G. Pilitsis
Christopher J. Nycz
Paulo A. W. G. Carvalho
Nirav Patel
Gregory S. Fischer
Marek Wartenberg
Radian Gondokaryono
Source :
IROS
Publication Year :
2017
Publisher :
IEEE, 2017.

Abstract

The use of magnetic resonance imaging (MRI) for guiding robotic surgical devices has shown great potential for performing precisely targeted and controlled interventions. To fully realize these benefits, devices must work safely within the tight confines of the MRI bore without negatively impacting image quality. Here we expand on previous work exploring MRI guided robots for neural interventions by presenting the mechanical design and assessment of a device for positioning, orienting, and inserting an interstitial ultrasound-based ablation probe. From our previous work we have added a 2 degree of freedom (DOF) needle driver for use with the aforementioned probe, revised the mechanical design to improve strength and function, and performed an evaluation of the mechanism’s accuracy and effect on MR image quality. The result of this work is a 7-DOF MRI robot capable of positioning a needle tip and orienting it’s axis with accuracy of 1.37 ± 0.06mm and 0.79° ± 0.41°, inserting it along it’s axis with an accuracy of 0.06 ± 0.07mm, and rotating it about it’s axis to an accuracy of 0.77° ± 1.31°. This was accomplished with no significant reduction in SNR caused by the robot’s presence in the MRI bore, ≤ 10.3% reduction in SNR from running the robot’s motors during a scan, and no visible paramagnetic artifacts.

Details

Database :
OpenAIRE
Journal :
2017 IEEE/RSJ International Conference on Intelligent Robots and Systems (IROS)
Accession number :
edsair.doi.dedup.....9baed05577518ce17b33f539737c1809