Back to Search Start Over

An easy-to-implement filter for separating photo-excited signals from topography in scanning tunneling microscopy

Authors :
Volker Rose
Daniel Rosenmann
Robert Winarski
Martin V. Holt
Saw-Wai Hla
Kangkang Wang
Source :
The Review of scientific instruments. 84(6)
Publication Year :
2013

Abstract

In order to achieve elemental and chemical sensitivity in scanning tunneling microscopy (STM), synchrotron x-rays have been applied to excite core-level electrons during tunneling. The x-ray photo-excitations result in tip currents that are superimposed onto conventional tunneling currents. While carrying important physical information, the varying x-ray induced currents can destabilize the feedback loop causing it to be unable to maintain a constant tunneling current, sometimes even causing the tip to retract fully or crash. In this paper, we report on an easy-to-implement filter circuit that can separate the x-ray induced currents from conventional tunneling currents, thereby allowing simultaneous measurements of topography and chemical contrasts. The filter and the schematic presented here can also be applied to other variants of light-assisted STM such as laser STM.