Back to Search Start Over

Development of a circulatory mock loop for biventricular device testing with various heart conditions

Authors :
David J. Horvath
Jamshid H. Karimov
Yuichiro Kado
Kiyotaka Fukamachi
Shengqiang Gao
Takuma Miyamoto
Source :
The International Journal of Artificial Organs. 43:600-605
Publication Year :
2020
Publisher :
SAGE Publications, 2020.

Abstract

This study aimed to evaluate a newly designed circulatory mock loop intended to model cardiac and circulatory hemodynamics for mechanical circulatory support device testing. The mock loop was built with dedicated ports suitable for attaching assist devices in various configurations. This biventricular mock loop uses two pneumatic pumps (Abiomed AB5000™, Danvers, MA, USA) driven by a dual-output driver (Thoratec Model 2600, Pleasanton, CA, USA). The drive pressures can be individually modified to simulate a healthy heart and left and/or right heart failure conditions, and variable compliance and fluid volume allow for additional customization. The loop output for a healthy heart was tested at 4.2 L/min with left and right atrial pressures of 1 and 5 mm Hg, respectively; a mean aortic pressure of 93 mm Hg; and pulmonary artery pressure of 17 mm Hg. Under conditions of left heart failure, these values were reduced to 2.1 L/min output, left atrial pressure = 28 mm Hg, right atrial pressure = 3 mm Hg, aortic pressure = 58 mm Hg, and pulmonary artery pressure = 35 mm Hg. Right heart failure resulted in the reverse balance: left atrial pressure = 0 mm Hg, right atrial pressure = 30 mm Hg, aortic pressure = 100 mm Hg, and pulmonary artery pressure = 13 mm Hg with a flow of 3.9 L/min. For biventricular heart failure, flow was decreased to 1.6 L/min, left atrial pressure = 13 mm Hg, right atrial pressure = 13 mm Hg, aortic pressure = 52 mm Hg, and pulmonary artery pressure = 18 mm Hg. This mock loop could become a reliable bench tool to simulate a range of heart failure conditions.

Details

ISSN :
17246040 and 03913988
Volume :
43
Database :
OpenAIRE
Journal :
The International Journal of Artificial Organs
Accession number :
edsair.doi.dedup.....a22616f8a29a3b7f4712662f05c24958