Thomson scattering on the large plasma device

Authors :
S. Ghazaryan
M. Kaloyan
W. Gekelman
Z. Lucky
S. Vincena
S. K. P. Tripathi
P. Pribyl
C. Niemann
Source :
Review of Scientific Instruments. 93:083514
Publication Year :
2022
Publisher :
AIP Publishing, 2022.

Abstract

We have developed a non-collective Thomson scattering diagnostic for measurements of electron density and temperature on the Large Plasma Device. A triple grating spectrometer with a tunable notch filter is used to discriminate the faint scattering signal from the stray light. In this paper, we describe the diagnostic and its calibration via Raman scattering and present the first measurements performed with the fully commissioned system. Depending on the discharge conditions, the measured densities and temperatures range from 4.0 × 1012 to 2.8 × 1013 cm−3 and from 1.2 to 6.8 eV, respectively. The variation of the measurement error with plasma parameters and discharges averaged is also discussed.

Details

ISSN :
10897623 and 00346748
Volume :
93
Database :
OpenAIRE
Journal :
Review of Scientific Instruments
Accession number :
edsair.doi.dedup.....a54f3becdaab9cfed6e234b534796246
Full Text :
https://doi.org/10.1063/5.0099172