Back to Search Start Over

Late recognition of Gore Helex atrial septal occluder fracture and migration

Authors :
Michael P. Cinquegrani
Harshal Patil
Noelle C. Garster
Lyle D. Joyce
Sudhi Tyagi
Michael H. Salinger
Paul J. Pearson
Source :
J Cardiol Cases
Publication Year :
2021
Publisher :
Japanese College of Cardiology, 2021.

Abstract

Patent foramen ovale (PFO) closure is indicated in cryptogenic stroke. Percutaneous PFO closure is both feasible and highly efficacious with low incidence of device-related complications. When complications occur, they are usually discovered within 6 weeks of device deployment. We describe the case of a partially embolised and fractured Gore Helex Septal Occluder device recognised nearly 9 years after placement requiring surgical explant.