Investigation of the Impact of Desorption Electrospray Ionization Sprayer Geometry on Its Performance in Imaging of Biological Tissue
- Source :
- Analytical Chemistry. 88:4808-4816
- Publication Year :
- 2016
- Publisher :
- American Chemical Society (ACS), 2016.
-
Abstract
- In this study, the impact of sprayer design and geometry on performance in desorption electrospray ionization mass spectrometry (DESI-MS) is assessed, as the sprayer is thought to be a major source of variability. Absolute intensity repeatability, spectral composition, and classification accuracy for biological tissues are considered. Marked differences in tissue analysis performance are seen between the commercially available and a lab-built sprayer. These are thought to be associated with the geometry of the solvent capillary and the resulting shape of the primary electrospray. Experiments with a sprayer with a fixed solvent capillary position show that capillary orientation has a crucial impact on tissue complex lipid signal and can lead to an almost complete loss of signal. Absolute intensity repeatability is compared for five lab-built sprayers using pork liver sections. Repeatability ranges from 1 to 224% for individual sprayers and peaks of different spectral abundance. Between sprayers, repeatability is 16%, 9%, 23%, and 34% for high, medium, low, and very low abundance peaks, respectively. To assess the impact of sprayer variability on tissue classification using multivariate statistical tools, nine human colorectal adenocarcinoma sections are analyzed with three lab-built sprayers, and classification accuracy for adenocarcinoma versus the surrounding stroma is assessed. It ranges from 80.7 to 94.5% between the three sprayers and is 86.5% overall. The presented results confirm that the sprayer setup needs to be closely controlled to obtain reliable data, and a new sprayer setup with a fixed solvent capillary geometry should be developed.
- Subjects :
- 0301 basic medicine
Spectrometry, Mass, Electrospray Ionization
Electrospray
Swine
Sprayer
Capillary action
0904 Chemical Engineering
Analytical chemistry
Geometry
Adenocarcinoma
01 natural sciences
Signal
Analytical Chemistry
03 medical and health sciences
0399 Other Chemical Sciences
Animals
Humans
Spectral composition
Desorption electrospray ionization
Chemistry
010401 analytical chemistry
Biological tissue
Repeatability
Lipids
Molecular Imaging
0104 chemical sciences
030104 developmental biology
Liver
Colorectal Neoplasms
0301 Analytical Chemistry
Subjects
Details
- ISSN :
- 15206882 and 00032700
- Volume :
- 88
- Database :
- OpenAIRE
- Journal :
- Analytical Chemistry
- Accession number :
- edsair.doi.dedup.....af90d64c62b438ab16ab14542cb0f39e
- Full Text :
- https://doi.org/10.1021/acs.analchem.6b00345