Back to Search Start Over

Novel App-Based Portable Spirometer for the Early Detection of COPD

Authors :
Chiung Zuei Chen
Cheng Yi Wang
Kang Yun Lee
Hen I. Lin
Jong Rung Tsai
Hao-Chien Wang
Chin Chou Wang
Meng-Chih Lin
Sheng Hao Lin
Diahn-Warng Perng
Shih Lung Cheng
Tsung Ming Yang
Wu-Huei Hsu
Ming Shian Lin
Ching Lung Liu
Ching Hsiung Lin
Tsai Yu Wang
Source :
Diagnostics, Diagnostics, Vol 11, Iss 785, p 785 (2021), Volume 11, Issue 5
Publication Year :
2021

Abstract

Chronic obstructive pulmonary disease (COPD) is preventable and treatable. However, many patients remain undiagnosed and untreated due to the underutilization or unavailability of spirometers. Accordingly, we used Spirobank Smart, an app-based spirometer, for facilitating the early detection of COPD in outpatient clinics. This prospective study recruited individuals who were at risk of COPD (i.e., with age of ≥40 years, ≥10 pack-years of smoking, and at least one respiratory symptoms) but had no previous COPD diagnosis. Eligible participants were examined with Spirobank Smart and then underwent confirmatory spirometry (performed using a diagnostic spirometer), regardless of their Spirobank Smart test results. COPD was defined and confirmed using the postbronchodilator forced expiratory volume in 1 s/forced vital capacity values of &lt<br />0.70 as measured by confirmatory spirometry. A total of 767 participants were enrolled and examined using Spirobank Smart<br />370 participants (94.3% men, mean age of 60.9 years and mean 42.6 pack-years of smoking) underwent confirmatory spirometry. Confirmatory spirometry identified COPD in 103 participants (27.8%). At the optimal cutoff point of 0.74 that was determined using Spirobank Smart for COPD diagnosis, the area under the receiver operating characteristic was 0.903 (95% confidence interval (CI) = 0.860–0.947). Multivariate logistic regression revealed that participants who have an FEV1/FVC ratio of &lt<br />74% that was determined using Spirobank Smart (odds ratio (OR) = 58.58, 95% CI = 27.29–125.75) and old age (OR = 3.23, 95% CI = 1.04–10.07 for 60 ≤ age &lt<br />65<br />OR = 5.82, 95% CI = 2.22–15.27 for age ≥ 65) had a higher risk of COPD. The Spirobank Smart is a simple and adequate tool for early COPD detection in outpatient clinics. Early diagnosis and appropriate therapy based on GOLD guidelines can positively influence respiratory symptoms and quality of life.

Details

ISSN :
20754418
Volume :
11
Issue :
5
Database :
OpenAIRE
Journal :
Diagnostics (Basel, Switzerland)
Accession number :