Effect of biphasic endocardial countershock on pacing threskolds in humans

Authors :
Rodney Horton
David J. Kessler
Robert Canby
Michael E. Jessen
Richard L. Page
Jose A. Joglar
Source :
The American Journal of Cardiology. 77:527-528
Publication Year :
1996
Publisher :
Elsevier BV, 1996.

Abstract

The use of non-thoracotomy endocardial implantable defibrillators with pacing capabilities has increased substantially over the past 2 years. This report demonstrates that the pacing threshold increases in some patients after endocardial defibrillation, and substantiates the practice of using maximal pacing output after endocardial defibrillation.