Back to Search
Start Over
Fast bound pool fraction mapping via steady‐state magnetization transfer saturation using single‐shot EPI
- Source :
- Magnetic Resonance in Medicine
- Publication Year :
- 2019
-
Abstract
- PURPOSE: To enable clinical applications of quantitative magnetization transfer (qMT) imaging by developing a fast method to map one of its fundamental model parameters, the bound pool fraction (BPF), in the human brain. THEORY AND METHODS: The theory of steady-state MT in the fast-exchange approximation is used to provide measurements of BPF, and bound pool transverse relaxation time ( T 2 B ). A sequence that allows sampling of the signal during steady-state MT saturation is used to perform BPF mapping with a 10-min-long fully echo planar imaging-based MRI protocol, including inversion recovery T1 mapping and B1 error mapping. The approach is applied in 6 healthy subjects and 1 multiple sclerosis patient, and validated against a single-slice full qMT reference acquisition. RESULTS: BPF measurements are in agreement with literature values using off-resonance MT, with average BPF of 0.114(0.100-0.128) in white matter and 0.068(0.054-0.085) in gray matter. Median voxel-wise percentage error compared with standard single slice qMT is 4.6%. Slope and intercept of linear regression between new and reference BPF are 0.83(0.81-0.85) and 0.013(0.11-0.16). Bland-Altman plot mean bias is 0.005. In the multiple sclerosis case, the BPF is sensitive to pathological changes in lesions. CONCLUSION: The method developed provides accurate BPF estimates and enables shorter scan time compared with currently available approaches, demonstrating the potential of bringing myelin sensitive measurement closer to the clinic.
- Subjects :
- Physics
Multiple Sclerosis
Echo-Planar Imaging
Healthy subjects
Single shot
Brain
Inversion recovery
030218 nuclear medicine & medical imaging
Scan time
03 medical and health sciences
0302 clinical medicine
Transverse Relaxation Time
Image Interpretation, Computer-Assisted
Linear regression
Humans
Radiology, Nuclear Medicine and imaging
Magnetization transfer
Saturation (magnetic)
Algorithm
Myelin Sheath
030217 neurology & neurosurgery
Subjects
Details
- ISSN :
- 07403194
- Database :
- OpenAIRE
- Journal :
- Magnetic Resonance in Medicine
- Accession number :
- edsair.doi.dedup.....b41967d13c90c2c40b0f980d85cf54ea
- Full Text :
- https://doi.org/10.1002/mrm.27792