Back to Search Start Over

Development of a Mass Sensitive Quartz Crystal Microbalance (QCM)-Based DNA Biosensor Using a 50 MHz Electronic Oscillator Circuit

Authors :
Bogdan Bucur
Gonzalo García-Martinez
Enrique Alonso Bustabad
Daniel Rose
Jose Farina
Chantal Compère
Claude Gabrielli
Mathieu Lazerges
Antonio Arnau Vives
Hubert Perrot
L. Rodriguez-Pardo
Departamento de Tecnología Electrónica
Universidade de Vigo
Laboratoire Interfaces et Systèmes Electrochimiques (LISE)
Université Pierre et Marie Curie - Paris 6 (UPMC)-Centre National de la Recherche Scientifique (CNRS)
Service Interfaces et Capteurs (IFREMER)
Institut Français de Recherche pour l'Exploitation de la Mer (IFREMER)
Department of Electronic Engineering [Universitat Politecnica de Valencia]
Universitat Politècnica de València (UPV)
Service Interfaces et capteurs (IC)
Source :
Sensors (1424-8220) (Mdpi Ag), 2011-08, Vol. 11, N. 8, P. 7656-7664, Sensors; Volume 11; Issue 8; Pages: 7656-7664, Sensors (Basel, Switzerland), RiuNet. Repositorio Institucional de la Universitat Politécnica de Valéncia, instname, Sensors, Sensors, MDPI, 2011, 11 (8), pp.7656-7664. ⟨10.3390/s110807656⟩, Sensors, Vol 11, Iss 8, Pp 7656-7664 (2011), Sensors, 2011, 11 (8), pp.7656-7664. ⟨10.3390/s110807656⟩
Publication Year :
2011
Publisher :
Mdpi Ag, 2011.

Abstract

This work deals with the design of a high sensitivity DNA sequence detector using a 50 MHz quartz crystal microbalance (QCM) electronic oscillator circuit. The oscillator circuitry is based on Miller topology, which is able to work in damping media. Calibration and experimental study of frequency noise are carried out, finding that the designed sensor has a resolution of 7.1 ng/cm 2 in dynamic conditions (with circulation of liquid). Then the oscillator is proved as DNA biosensor. Results show that the system is able to detect the presence of complementary target DNAs in a solution with high selectivity and sensitivity. DNA target concentrations higher of 50 ng/mL can be detected. © 2011 by the authors; licensee MDPI, Basel, Switzerland.<br />ANR PRECODD (HAB SEACHIP) and Erasmus Program (European Union) are gratefully acknowledged for financially supporting this research.

Details

Language :
English
ISSN :
14248220
Database :
OpenAIRE
Journal :
Sensors (1424-8220) (Mdpi Ag), 2011-08, Vol. 11, N. 8, P. 7656-7664, Sensors; Volume 11; Issue 8; Pages: 7656-7664, Sensors (Basel, Switzerland), RiuNet. Repositorio Institucional de la Universitat Politécnica de Valéncia, instname, Sensors, Sensors, MDPI, 2011, 11 (8), pp.7656-7664. ⟨10.3390/s110807656⟩, Sensors, Vol 11, Iss 8, Pp 7656-7664 (2011), Sensors, 2011, 11 (8), pp.7656-7664. ⟨10.3390/s110807656⟩
Accession number :
edsair.doi.dedup.....b70afb20fdcc98dee41f4d28a0663fef
Full Text :
https://doi.org/10.3390/s110807656⟩