Back to Search Start Over

Effect of refraction on optical microscopic measurement of internal blood-vessel diameter and its correction

Authors :
Eric Q. Chen
Chan F. Lam
Ammafi Periasamy
Source :
Journal of microscopy. 164(Pt 3)
Publication Year :
1991

Abstract

SUMMARY In optical microscopic measurement of internal blood-vessel diameters, the effect of refraction must be taken into account to ensure accuracy of the result. This effect is discussed and an analytical correction formula derived. Phantom blood vessels with known internal and external diameters were used to test the validity of the correction formula. The errors obtained prior to correction were reduced significantly after correction.

Details

ISSN :
00222720
Volume :
164
Issue :
Pt 3
Database :
OpenAIRE
Journal :
Journal of microscopy
Accession number :
edsair.doi.dedup.....b7f69e36b86e8d3d9dd501765b210f88