Abstract

To evaluate the usefulness of the advanced monoenergetic imaging (AMI) reconstruction technique for dual-energy computed tomography to evaluate endoleaks after endovascular stent-graft placement. Ninety-five dual-phase (early and delayed phases) enhanced CT examinations were performed for 60 patients who underwent endovascular stent-graft placement. AM images were reconstructed at 40 keV and compared with the standard 120-kVp images (SI). The signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR) of the aorta and endoleak were measured. Two radiologists subjectively assessed endoleak delineation and contrast enhancement conditions using a 5-point Likert scale (1: poor—5: excellent). Mean SNRs of the aorta were higher by AMI (early; 34.7 ± 10.2 [SD], delay; 11.4 ± 3.2) than by SI (early; 23.1 ± 6.3, delay; 8.6 ± 2.2) (P