Back to Search
Start Over
Evaluation of Matrix-assisted laser desorption/ionization Time-of flight Mass spectrometry (MALDI TOF MS) and VITEK 2 in routine microbial identification
- Source :
- Ghana Medical Journal; Vol. 55 No. 4 (2021); 308-310
- Publication Year :
- 2021
- Publisher :
- African Journals Online (AJOL), 2021.
-
Abstract
- Background: Microbial Identification was done by phenotypic methods. VITEK-2 and Matrix-assisted laser desorption ionization-time of flight mass spectrometry (MALDI-TOF MS) are now being increasingly used in laboratories.Objectives: To compare and evaluate the usefulness of MALDI-TOF MS and VITEK-2 in routine microbial identification.Methods: The performances of MALDI-TOF MS and VITEK 2 were compared for identifying microorganisms.Results: MALDI- TOF MS and VITEK-2 correctly identified 96 % (96/100) and 97% (97/100) of the isolates upto the genus level.Conclusion: MALDI TOF MS and VITEK -2 gave comparable identification and error rates. The rapid reduction in turnaround time with MALDI TOF is a significant game-changer in the field of clinical microbiology
Details
- ISSN :
- 00169560
- Volume :
- 55
- Database :
- OpenAIRE
- Journal :
- Ghana Medical Journal
- Accession number :
- edsair.doi.dedup.....bae0449dd4d27a8f88d779ca1c6dcf61
- Full Text :
- https://doi.org/10.4314/gmj.v55i4.12