Efficacy of 3D transoesophageal echocardiography for transcatheter device closure of atrial septal defect without balloon sizing

Authors :
Duk-Woo Park
Jung-Min Ahn
Jae-Kwan Song
Jong-Min Song
Jaeseok Bae
Ran Heo
Duk-Hyun Kang
Dae-Hee Kim
Sahmin Lee
Min Soo Cho
Jeong Yoon Jang
Jung Ae Hong
Source :
European Heart Journal - Cardiovascular Imaging. 19:684-689
Publication Year :
2017
Publisher :
Oxford University Press (OUP), 2017.

Abstract

Aims Using balloon sizing to determine device size may cause complications and increase procedure time in performing transcatheter closure of atrial septal defect (ASD). We aimed to validate the clinical utility of a formula using measurements from 3D transoesophageal echocardiography (TOE) images in performing the procedure without balloon sizing. Methods and results We enrolled 248 consecutive patients with ASD in a prospective registry. In the first tier (n = 53), we determined the device size before the procedure using our formula and performed balloon sizing during the procedure to verify our decision. In the second tier (n = 195), the procedure was performed without balloon sizing. In the first tier, the estimated device size correlated well with the device size finally implanted (R = 0.961, P