Diagnostic accuracy of fluorescence flow-cytometry technology using Sysmex XN-31 for imported malaria in a non-endemic setting

Authors :
Stéphane Picot
Thomas Perpoint
Christian Chidiac
Alain Sigal
Etienne Javouhey
Yves Gillet
Laurent Jacquin
Marion Douplat
Karim Tazarourte
Laurent Argaud
Martine Wallon
Charline Miossec
Guillaume Bonnot
Anne-Lise Bienvenu
HCL Groupement Hospitalier Nord [Lyon]
Institut de Chimie et Biochimie Moléculaires et Supramoléculaires (ICBMS)
Université Claude Bernard Lyon 1 (UCBL)
Université de Lyon-Université de Lyon-École Supérieure de Chimie Physique Électronique de Lyon (CPE)-Institut National des Sciences Appliquées de Lyon (INSA Lyon)
Université de Lyon-Institut National des Sciences Appliquées (INSA)-Institut National des Sciences Appliquées (INSA)-Institut de Chimie du CNRS (INC)-Centre National de la Recherche Scientifique (CNRS)
Hôpital de la Croix-Rousse [CHU - HCL]
Hospices Civils de Lyon (HCL)
Centre International de Recherche en Infectiologie (CIRI)
École normale supérieure de Lyon (ENS de Lyon)-Université Claude Bernard Lyon 1 (UCBL)
Université de Lyon-Université de Lyon-Université Jean Monnet - Saint-Étienne (UJM)-Institut National de la Santé et de la Recherche Médicale (INSERM)-Centre National de la Recherche Scientifique (CNRS)
Hôpital Femme Mère Enfant [CHU - HCL] (HFME)
Université de Lyon
Hôpital Edouard Herriot [CHU - HCL]
Centre Hospitalier Lyon Sud [CHU - HCL] (CHLS)
Health Service and Performance Research (HESPER)
Université de Lyon-Université de Lyon
Université de Lyon-Université de Lyon-Institut National des Sciences Appliquées de Lyon (INSA Lyon)
Université de Lyon-Institut National des Sciences Appliquées (INSA)-Institut National des Sciences Appliquées (INSA)-Institut de Chimie du CNRS (INC)-École Supérieure Chimie Physique Électronique de Lyon-Centre National de la Recherche Scientifique (CNRS)
Centre International de Recherche en Infectiologie - UMR (CIRI)
Institut National de la Santé et de la Recherche Médicale (INSERM)-École normale supérieure - Lyon (ENS Lyon)-Université Claude Bernard Lyon 1 (UCBL)
Université de Lyon-Université de Lyon-Centre National de la Recherche Scientifique (CNRS)
École normale supérieure - Lyon (ENS Lyon)-Université Claude Bernard Lyon 1 (UCBL)
Université de Lyon-Université de Lyon-Institut National de la Santé et de la Recherche Médicale (INSERM)-Centre National de la Recherche Scientifique (CNRS)
Sciences, EDP
Source :
Parasite, Parasite, 2022, 29, pp.31. ⟨10.1051/parasite/2022031⟩, Parasite, EDP Sciences, 2022, 29, pp.31. ⟨10.1051/parasite/2022031⟩
Publication Year :
2022
Publisher :
HAL CCSD, 2022.

Abstract

Malaria diagnosis based on microscopy is impaired by the gradual disappearance of experienced microscopists in non-endemic areas. Aside from the conventional diagnostic methods, fluorescence flow cytometry technology using Sysmex XN-31, an automated haematology analyser, has been registered to support malaria diagnosis. The aim of this prospective, monocentric, non-interventional study was to evaluate the diagnostic accuracy of the XN-31 for the initial diagnosis or follow-up of imported malaria cases compared to the reference malaria tests including microscopy, loop mediated isothermal amplification, and rapid diagnostic tests. Over a one-year period, 357 blood samples were analysed, including 248 negative and 109 positive malaria samples. Compared to microscopy, XN-31 showed sensitivity of 100% (95% CI: 97.13-100) and specificity of 98.39% (95% CI: 95.56-100) for the initial diagnosis of imported malaria cases. Moreover, it provided accurate species identification asfalciparumor non-falciparumand parasitaemia determination in a very short time compared to other methods. We also demonstrated that XN-31 was a reliable method for patient follow-up on days 3, 7, and 28. Malaria diagnosis can be improved in non-endemic areas by the use of dedicated haematology analysers coupled with standard microscopy or other methods in development, such as artificial intelligence for blood slide reading. Given that XN-31 provided an accurate diagnosis in 1 min, it may reduce the time interval before treatment and thus improve the outcome of patient who have malaria.Précision diagnostique de la technique de cytométrie de flux en fluorescence utilisant le Sysmex XN-31 pour le paludisme d’importation en zone non endémique.Le diagnostic du paludisme basé sur la microscopie est rendu difficile par la disparition progressive des microscopistes expérimentés en zone non-endémique. À côté des méthodes conventionnelles, la technique de cytométrie de flux en fluorescence utilisant le Sysmex XN-31, un automate d’analyse hématologique, a été enregistrée pour participer au diagnostic du paludisme. L’objectif de cette étude prospective, monocentrique et non-interventionelle était d’évaluer la précision diagnostique du XN-31 pour le diagnostic initial et le suivi des cas de paludisme d’importation en comparaison des tests de référence dont la microscopie, l’amplification isothermale en boucle, et des tests de diagnostic rapide. Durant une période d’un an, 357 échantillons de sang ont été analysés, dont 248 négatifs et 109 positifs pour le paludisme. En comparaison de la microscopie, le XN-31 a montré une sensibilité de 100 % (95 % CI : 97.13-100) et une spécificité de 98.39 % (95 % CI : 95.56-100) pour le diagnostic initial des cas de paludisme d’importation. De plus, l’identification des espècesfalciparumet non-falciparumainsi que la parasitémie ont été précises dans un temps très court en comparaison des autres méthodes. Nous avons aussi démontré que le XN-31 était une méthode fiable pour le suivi des patients à J3, J7 et J28. Le diagnostic du paludisme peut être amélioré en zone non-endémique par l’utilisation d’automates d’hématologie spécialisés, associés à la microscopie standard ou d’autres méthodes en développement telle que l’intelligence artificielle appliquée à la lecture des lames de sang. Dans la mesure où le XN-31 produit un diagnostic précis en une minute, cela peut réduire le délai avant le traitement et donc améliorer l’issue pour les patients souffrant de paludisme.