The purpose of this study was to describe the use of the VIPER software for patient-specific quality assurance (PSQA) of single-isocenter multitarget (SIMT) stereotactic radiosurgery (SRS) plans.Twenty clinical of intensity-modulated (IMRT) SIMT SRS plans were reviewed. A total of 88 brain metastases were included. Number of lesions per plan and their individual volumes ranged from 2 to 35 and from 0.03 to 32.8 cmThe averages (± 1 SD) of the 3D GPRs over the 20 SRS plans were: 99.9 ± 0.2%, 99.7 ± 0.3%, 99.6 ± 0.5%, 99.3 ± 0.9%,99.1 ± 1.6%, 99.0 ± 1.6%, and 98.5 ± 3.3%, for dose thresholds of 10%, 20%, 30%, 50%, 70%, 80% and 90% respectively.This work shows the feasibility of the VIPER software for PSQA of SIMT SRS plans, being a reliable alternative to commercially available 2D detector arrays.