Minimum acceptable sensitivity of intraoperative gamma probes used for sentinel lymph node detection in melanoma patients

Authors :
Marco Brambilla
Roberta Matheoud
S. Valzano
R. Giorgione
E. Colombo
Gian Mauro Sacchetti
Source :
Physica Medica. 30:822-826
Publication Year :
2014
Publisher :
Elsevier BV, 2014.

Abstract

The aim of this study was to determine the suspension level for the sensitivity of an intraoperative scintillation gamma probe in the detection of the sentinel lymph node (SLN) in melanoma patients. Thirty-eight consecutive patients with melanoma were enrolled in the study during a 12-month period and underwent lymphatic scintigraphy after the peritumoral intradermal administration of about 14 MBq of 99m Tc-nanocolloids. The SLNs were successfully removed during the surgical intervention about 4 h later. To identify and localize the SLN, a scintillation NaI(Tl) collimated probe was used. Predictably, the probe sensitivity decreased as the photopeak energy window was progressively narrowed, from 6.9 ± 0.7 counts per second (cps)/kBq (designated as the ‘optimum,’ or ‘OPT,’ sensitivity) to 2.5 ± 0.3 cps/kBq (LOW sensitivity) and to 1.4 ± 0.2 cps/kBq (VLOW sensitivity). Maximum lymph node count rates (cps) were determined for the foregoing energy windows prior to skin incision (PRE OPT , PRE LOW , PRE VLOW , respectively) and in vivo after incision (INV OPT , INV LOW , INV VLOW ). Forty-three SLNs were removed with a mean source-to-detector distance of 46 ± 24 mm (min 12 mm, max 92 mm). Four SLNs could not have been detected using PRE LOW. This figure would have decreased to 34, with nine undetectable lymph nodes, with PRE VLOW. One SLN could not have been identified using INV LOW and four could not have be identified using INV VLOW. In the clinical scenario of SLN detection in melanoma patients, a system sensitivity of 2.5 cps/kBq represents a suspension level, that is, a level under which the equipment must be suspended from clinical use and the poor performance must be investigated.

Details

ISSN :
11201797
Volume :
30
Database :
OpenAIRE
Journal :
Physica Medica
Accession number :
edsair.doi.dedup.....bf53f502894c113054f886fd0650812c
Full Text :
https://doi.org/10.1016/j.ejmp.2014.03.006