Back to Search
Start Over
A comparison of two pencil beam scanning treatment planning systems for proton therapy
- Source :
- Journal of Applied Clinical Medical Physics
- Publication Year :
- 2017
-
Abstract
- Objective Analytical dose calculation algorithms for Eclipse and Raystation treatment planning systems (TPS), as well as a Raystation Monte Carlo model are compared to corresponding measured point doses. Method The TPS were modeled with the same beam data acquired during commissioning. Thirty-five typical plans were made with each planning system, 31 without range shifter and four with a 5 cm range shifter. Point doses in these planes were compared to measured doses. Results The mean percentage difference for all plans between Raystation and Eclipse were 1.51 ± 1.99%. The mean percentage difference for all plans between TPS models and measured values are −2.06 ± 1.48% for Raystation pencil beam (PB), −0.59 ± 1.71% for Eclipse and −1.69 ± 1.11% for Raystation monte carlo (MC). The distribution for the patient plans were similar for Eclipse and Raystation MC with a P-value of 0.59 for a two tailed unpaired t-test and significantly different from the Raystation PB model with P = 0.0013 between Raystation MC and PB. All three models faired markedly better if plans with a 5 cm range shifter were ignored. Plan comparisons with a 5 cm range shifter give differences between Raystation and Eclipse of 3.77 ± 1.82%. The mean percentage difference for 5 cm range shifter plans between TPS models and measured values are −3.89 ± 2.79% for Raystation PB, −0.25 ± 3.85% for Eclipse and 1.55 ± 1.95% for Raystation MC. Conclusion Both Eclipse and Raystation PB TPS are not always accurate within ±3% for a 5 cm range shifters or for small targets. This was improved with the Raystation MC model. The point dose calculations of Eclipse, Raystation PB, and Raystation MC compare within ±3% to measured doses for the other scenarios tested.
- Subjects :
- treatment planning
Dose calculation
030218 nuclear medicine & medical imaging
03 medical and health sciences
0302 clinical medicine
Neoplasms
Proton Therapy
Humans
Radiation Oncology Physics
Radiology, Nuclear Medicine and imaging
Radiation treatment planning
Pencil-beam scanning
Instrumentation
Proton therapy
87.55.d
Eclipse
Physics
Radiation
business.industry
Phantoms, Imaging
Radiotherapy Planning, Computer-Assisted
87.55.k
Radiotherapy Dosage
Monte Carlo methods
030220 oncology & carcinogenesis
Percentage difference
Radiotherapy, Intensity-Modulated
Nuclear medicine
business
Monte Carlo Method
Algorithms
Subjects
Details
- ISSN :
- 15269914
- Volume :
- 19
- Issue :
- 1
- Database :
- OpenAIRE
- Journal :
- Journal of applied clinical medical physics
- Accession number :
- edsair.doi.dedup.....c0999ba6722042cac230706990f5c932