- Source :
- Pacing and clinical electrophysiology : PACE. 42(8)
- Publication Year :
- 2018
-
Abstract
- Background While electromagnetic interference is widely recognized as a cause of noise detected by cardiac implanted electronic devices (CIEDs), we report on two cases of inappropriate activation of magnet response mode due to magnets in continuous positive airway pressure (CPAP) masks. Methods Two patients with St. Jude Medical dual-chamber pacemakers (Assurity DR 2240; St. Jude Medical, St. Paul, MN, USA) were discovered on routine outpatient device interrogation to have magnet response events that correlated with nightly use of their CPAP masks. A magnet response event was reproduced by positioning one CPAP mask over the patient's pulse generator site. The magnetic flux density was measured using a Gauss meter. Results It was found that their CPAP masks had magnetic clips that were used to secure the headgears. These appeared to be transiently positioned within sufficient proximity of the patients' CIEDs to activate magnet response events during sleep. The measured magnetic field density adjacent to a mask magnet was 136 Gauss. Conclusion Clinicians managing CPAPs and CIEDs should be aware of magnetic interference with the potential for inappropriate activation of magnet response mode. CPAP masks that are secured with magnetic clips should be avoided in patients with CIEDs, since this can result in inappropriate asynchronous pacing in pacemakers or potential failure to deliver life-saving tachytherapies in patients with implanted cardioverter-defibrillators.
- Subjects :
- medicine.medical_specialty
Pacemaker, Artificial
medicine.medical_treatment
030204 cardiovascular system & hematology
03 medical and health sciences
0302 clinical medicine
Internal medicine
medicine
Humans
In patient
030212 general & internal medicine
Continuous positive airway pressure
Aged
Aged, 80 and over
Continuous Positive Airway Pressure
business.industry
Masks
Cardiac implanted
General Medicine
equipment and supplies
Magnet
Cardiology
Magnets
Equipment Failure
Cardiology and Cardiovascular Medicine
business
human activities
Subjects
Details
- ISSN :
- 15408159
- Volume :
- 42
- Issue :
- 8
- Database :
- OpenAIRE
- Journal :
- Pacing and clinical electrophysiology : PACE
- Accession number :
- edsair.doi.dedup.....c3b5a3d476cd7dd05539078b6558bdf2