Dosimetric effect of Elekta Beam Modulator micromultileaf in three-dimensional conformal radiotherapy and intensity-modulated radiotherapy for prostate cancer

Authors :
Alessandra Murgia
Elisabetta Ponti
Barbara Tolu
Alessandra Carosi
Riccardo Santoni
Daniela di Cristino
Gianluca Ingrosso
Source :
Medical Dosimetry. 39:180-184
Publication Year :
2014
Publisher :
Elsevier BV, 2014.

Abstract

The purpose of this study is to analyze the dosimetric effect of Elekta Beam Modulator in 3-dimensional conformal radiation therapy (3DCRT) and in intensity-modulated radiation therapy (IMRT) for localized prostate cancer. We compared treatment plans developed with 2 different Elekta multileaf collimators (MLC): Beam Modulator micro-MLC (mMLC) (4-mm leaf width at the isocenter) and standard MLC (10-mm leaf width at the isocenter). The comparison was performed for 15 patients with localized prostate cancer in 3DCRT and IMRT delivery; a total of 60 treatment plans were processed. The dose-volume histograms were used to provide the quantitative comparison between plans. In particular, we analyzed differences between rectum and bladder sparing in terms of a set of appropriate Vx (percentage of organ at risk [OAR] volume receiving the x dose) and differences between target conformity and coverage in terms of coverage factor and conformation number. Our analysis demonstrates that in 3DCRT there is an advantage in the use of Elekta Beam Modulator mMLC in terms of organ sparing; in particular, a significant decrease in rectal V60 and V50 (p = 0.001) and in bladder V70 and V65 (p = 0.007 and 0.002, respectively) was found. Moreover, a better target dose conformity was obtained (p = 0.002). IMRT plans comparison demonstrated no significant differences between the use of the 4 or 10-mm MLCs. Our analysis shows that in 3DCRT the use of the Elekta Beam Modulator mMLC gives a gain in target conformity and in OARs dose sparing whereas in IMRT plans there is no advantage.

Details

ISSN :
09583947
Volume :
39
Database :
OpenAIRE
Journal :
Medical Dosimetry
Accession number :
edsair.doi.dedup.....c3fb779a0fe4e378fca56622659edbe1
Full Text :
https://doi.org/10.1016/j.meddos.2013.12.006