Back to Search
Start Over
Feasibility and sensitivity study of helical tomotherapy for dose painting plans
- Source :
- Acta Oncologica. 49:991-996
- Publication Year :
- 2010
- Publisher :
- Informa UK Limited, 2010.
-
Abstract
- Important limitations for dose painting are due to treatment planning and delivery constraints. The purpose of this study was to develop a methodology for creating voxel-based dose painting plans that are deliverable using the clinical TomoTherapy Hi-Art II treatment planning system (TPS). Material and methods. Uptake data from a head and neck patient who underwent a [(61)Cu]Cu-ATSM (hypoxia surrogate) PET/CT scan was retrospectively extracted for planning. Non-uniform voxel-based prescriptions were converted to structured-based prescriptions for compatibility with the Hi-Art II TPS. Optimized plans were generated by varying parameters such as dose level, structure importance, prescription point normalization, DVH volume, min/max dose, and dose penalty. Delivery parameters such as pitch, jaw width and modulation factor were also varied. Isodose distributions, quality volume histograms and planning target volume percentage receiving planned dose within 5% of the prescription (Q(0.95-1.05)) were used to evaluate plan conformity. Results. In general, the conformity of treatment plans to dose prescriptions was found to be adequate for delivery of dose painting plans. The conformity was better as the dose levels increased from three to nine levels (Q(0.95-1.05): 69% to 93%), jaw decreased in width from 5.0cm to 1.05cm (Q(0.95-1.05): 81% to 93%), and modulation factor increased up to 2.0 (Q(0.95-1.05): 36% to 92%). The conformity was invariant to changes in pitch. Plan conformity decreased as the prescription DVH constraint (Q(0.95-1.05): 93% vs. 89%) or the normalization point (Q(0.95-1.05): 93% vs. 90%) deviated from the means. Conclusion. This investigation demonstrated the ability of the Hi-Art II TPS to create voxel-based dose painting plans. Results indicated that agreement in prescription dose and planned dose distributions for all plans were sensitive to physical delivery parameter changes in jaw width and modulation factors, but insensitive to changes in pitch. Tight constraints on target structures also resulted in decreased plan conformity while under a relaxed set of optimization parameters, plan conformity was increased.
- Subjects :
- Thiosemicarbazones
medicine.medical_specialty
medicine.medical_treatment
computer.software_genre
Sensitivity and Specificity
Article
Tomotherapy
Planned Dose
Coordination Complexes
Voxel
Dose painting
Organometallic Compounds
Humans
Medicine
Radiology, Nuclear Medicine and imaging
Sensitivity (control systems)
Radiation treatment planning
Head and neck
medicine.diagnostic_test
business.industry
Radiotherapy Planning, Computer-Assisted
Radiotherapy Dosage
Hematology
General Medicine
Tumor Burden
Copper Radioisotopes
Oncology
Head and Neck Neoplasms
Positron emission tomography
Positron-Emission Tomography
Carcinoma, Squamous Cell
Feasibility Studies
Radiology
Radiotherapy, Conformal
business
Nuclear medicine
Tomography, Spiral Computed
computer
Subjects
Details
- ISSN :
- 1651226X and 0284186X
- Volume :
- 49
- Database :
- OpenAIRE
- Journal :
- Acta Oncologica
- Accession number :
- edsair.doi.dedup.....ce29e6120414aab415c5b1c7df0a1e2a