Back to Search
Start Over
Conventional and Molecular Methods for Verification of Results Obtained with BacT/Alert Nonvent Blood Culture Bottles
- Source :
- Journal of Clinical Microbiology. 41:1266-1269
- Publication Year :
- 2003
- Publisher :
- American Society for Microbiology, 2003.
-
Abstract
- A strategy comparing molecular and conventional methods for verification of the BacT/Alert nonvent blood culture bottles (Organon Teknika, Durham, N.C.) was performed with seeded isolates. The bottles were evaluated with 12 common organisms from bloodstream infections. Overall, the bottles were equivalent as determined by conventional and molecular methods.
Details
- ISSN :
- 1098660X and 00951137
- Volume :
- 41
- Database :
- OpenAIRE
- Journal :
- Journal of Clinical Microbiology
- Accession number :
- edsair.doi.dedup.....ce5fa130e530fd748b16860b11ffdaa6
- Full Text :
- https://doi.org/10.1128/jcm.41.3.1266-1269.2003