Back to Search
Start Over
Compact and versatile neutron imaging detector with sub 4 mu m spatial resolution based on a single crystal thin film scintillator
- Source :
- 'Optics Express ', vol: 30, pages: 14461-14477 (2022)
- Publication Year :
- 2022
-
Abstract
- A large and increasing number of scientific domains pushes for high neutron imaging resolution achieved in reasonable times. Here we present the principle, design and performance of a detector based on infinity corrected optics combined with a crystalline Gd3Ga5O12 : Eu scintillator, which provides an isotropic sub-4 µm true resolution. The exposure times are only of a few minutes per image. This is made possible also by the uniquely intense cold neutron flux available at the imaging beamline NeXT-Grenoble. These comparatively rapid acquisitions are compatible with multiple high quality tomographic acquisitions, opening new venues for in-operando testing, as briefly exemplified here.
- Subjects :
- high neutron imaging resolution
detector
Atomic and Molecular Physics, and Optics
Subjects
Details
- Database :
- OpenAIRE
- Journal :
- 'Optics Express ', vol: 30, pages: 14461-14477 (2022)
- Accession number :
- edsair.doi.dedup.....cf57378283ecb95daad442c55fd2cea4