Back to Search
Start Over
Neurologic Events With a Totally Implantable Left Ventricular Assist Device: European LionHeart Clinical Utility Baseline Study (CUBS)
Neurologic Events With a Totally Implantable Left Ventricular Assist Device: European LionHeart Clinical Utility Baseline Study (CUBS)
- Source :
- The Journal of Heart and Lung Transplantation. 26:1-8
- Publication Year :
- 2007
- Publisher :
- Elsevier BV, 2007.
-
Abstract
- Background Neurologic events such as thromboembolic and hemorrhagic strokes are common complications of mechanical circulatory support. We report the neurologic events observed in patients treated for end-stage heart failure with the implantable, pulsatile LionHeart left ventricular assist device (LVAD). This sub-study was part of the LionHeart European Clinical Utility Baseline Study (CUBS). Methods Twenty-three male patients were implanted with the LionHeart LVAD in a non-randomized, observational study. Neurologic events were classified into three categories: (1) transient ischemic attacks (TIAs); (2) strokes, including cerebrovascular accidents (CVAs) and intracranial bleeding (ICB); and (3) "other," including hypoperfusion, coma and brain death. Neurologic injuries were also categorized as transient/reversible or permanent/disabling. Results Thirteen of 23 patients (57%) had a total of 30 neurologic events. Eight patients (35%) had 18 TIAs. Eight patients (35%) also had a stroke, either CVA ( n = 5, 22%) or ICB ( n = 3, 13%), and 5 of these patients (22%) also had 12 TIAs. Three patients (13%) had 4 "other" neurologic events. Ten patients (43%) had transient/reversible neurologic deficits and 10 (43%) had permanent/disabling events. One patient (4%) had intracranial bleeding as a primary cause of death (anti-coagulation–related hemorrhage). The combined incidence of neurologic events was 1.37 events/patient-year. The incidences of transient and permanent events were 0.91 and 0.46 event/patient-year, respectively. Conclusions Neurologic events caused morbidity in the CUBS trial, with infrequent mortality. These results are similar to previous experiences with destination therapy and underscore the need for improvements in LVAD design, patient selection and patient management to reduce the incidence of neurologic events.
- Subjects :
- Male
Pulmonary and Respiratory Medicine
medicine.medical_specialty
medicine.medical_treatment
law.invention
Risk Factors
law
Artificial heart
Internal medicine
medicine
Humans
Prospective Studies
Prospective cohort study
Aged
Cause of death
Heart Failure
Coma
Transplantation
business.industry
Incidence
Middle Aged
Prognosis
medicine.disease
Surgery
Stroke
Clinical trial
Ventricular assist device
Heart failure
Cardiology
France
Heart-Assist Devices
medicine.symptom
Cardiology and Cardiovascular Medicine
business
Follow-Up Studies
Destination therapy
Subjects
Details
- ISSN :
- 10532498
- Volume :
- 26
- Database :
- OpenAIRE
- Journal :
- The Journal of Heart and Lung Transplantation
- Accession number :
- edsair.doi.dedup.....cf74cda924e8bb76979905c0c401e5d0
- Full Text :
- https://doi.org/10.1016/j.healun.2006.10.007