Sensitive determination of malondialdehyde in rat prostate by high performance liquid chromatography with fluorescence detection

Authors :
Xiuli Dong
Xiangming Chen
Jiayuan Tang
Source :
Scientific Reports, Vol 10, Iss 1, Pp 1-7 (2020), Scientific Reports
Publication Year :
2020
Publisher :
Nature Publishing Group, 2020.

Abstract

An excellent pre-column fluorescent derivatization reagent N-acetylhydrazine acridone for the quantitative determination of malondialdehyde was synthesized. Malondialdehyde was derivatized at 80 °C for 30 min in the presence of trichloroacetic acid. The separation of the derivative was performed on an Agilent ZORBAX SB-C18 column in conjunction with gradient elution. The excitation and emission wavelengths were 370 nm and 420 nm, respectively. The developed method demonstrated good linear relationship in the range of 0.02 pmol to 2.5 pmol (r = 0.9998). The calculated limit of detection and limit of quantification were 2.5 fmol and 8.3 fmol, respectively. The analytical precisions of the method were in the range of 1.36–2.27% (intra-day) and 2.36–3.92% (inter-day) respectively. The method was sensitive, specific and simple. It was successfully implemented to analysis the malondialdehyde in rat prostate.

Details

Language :
English
ISSN :
20452322
Volume :
10
Issue :
1
Database :
OpenAIRE
Journal :
Scientific Reports
Accession number :
edsair.doi.dedup.....d1e3060ac08779357f2ca52a3948eb62