Reliability and Influence on Decision Making of fully-automated vs. semi-automated Software Packages for Procedural Planning in TAVI

Authors :
Markus Kofler
Jörg Kempfert
Semih Buz
Natalia Solowjowa
Volkmar Falk
Maximilian Y. Emmert
Christof Stamm
Axel Unbehaun
Matteo Montagner
Alexander Meyer
S. Sündermann
Christoph Klein
University of Zurich
Kofler, Markus
Source :
Scientific Reports, 10 (1), Scientific Reports, Scientific Reports, Vol 10, Iss 1, Pp 1-8 (2020)
Publication Year :
2020
Publisher :
Nature Publishing Group, 2020.

Abstract

Precise procedural planning is crucial to achieve excellent results in patients undergoing Transcatheter aortic valve implantation (TAVI). The aim of this study was to compare the semi-automated 3mensio (3 m) software to the fully-automated HeartNavigator3 (HN) software. We randomly selected 100 patients from our in-house TAVI-registry and compared aortic annulus and perimeter as well as coronary distances between 3m-measurements and post-hoc HN-measurements. Finally, we retrospectively simulated prosthesis choice based on HN-measurements and analyzed the differences compared to routinely used 3 m based strategy. We observed significant differences between the two software packages regarding area (3 m 464 ± 88 mm², HN 482 ± 96 mm², p<br />Scientific Reports, 10 (1)<br />ISSN:2045-2322