Abstract

ObjectivesSurgical management of tricuspid valve regurgitation mainly consists of tricuspid valve annuloplasty, usually performed with implantation of a rigid ring or a flexible band.MethodsWe performed a retrospective analysis on 820 patients who underwent tricuspid valve repair between March 2002 and July 2009 with either a flexible Cosgrove-Edwards band (n = 415; Edwards Lifesciences LLC, Irvine, Calif) or a rigid Carpentier-Edwards Classic annuloplasty ring (n = 405; Edwards Lifesciences). Mean patient age was 69.2 ± 9.5 years, 54.1% were female, and average logistic EuroSCORE was 13.3% ± 12.5%. Concomitant procedures were performed in 94.6% of patients (mitral valve surgery, 80.6%; aortic valve surgery, 28.2%; coronary artery bypass grafting, 24.5%; atrial fibrillation ablation, 44.5%). One fifth of the operations were reoperative procedures. Follow-up was 94% complete, with mean duration of 21.0 ± 19.0 months.ResultsThirty-day mortality was 10.1% (Cosgrove-Edwards, 11.9%; Carpentier-Edwards, 8.4%), and 5-year survival was 62.4% (Carpentier-Edwards, 64.7%; Cosgrove-Edwards, 60.3%). Postoperative echocardiography showed significant improvement in tricuspid valve function, with reduction in tricuspid regurgitation grade from 2.3 ± 0.7 to 0.7 ± 0.7, and no differences between groups. Use of a Carpentier-Edwards ring, however, was associated with significantly higher risk of dehiscence (Carpentier-Edwards, 8.7%; Cosgrove-Edwards, 0.9%; P

Details

ISSN :
00225223
Volume :
143
Database :
OpenAIRE
Journal :
The Journal of Thoracic and Cardiovascular Surgery
Accession number :
edsair.doi.dedup.....d5d03a90740481cb312363ac66efd118