Back to Search
Start Over
Performance of A Computerized CCU Arrhythmia Monitor System
- Source :
- Journal of Clinical Engineering. 7:201-210
- Publication Year :
- 1982
- Publisher :
- Ovid Technologies (Wolters Kluwer Health), 1982.
-
Abstract
- Experience gained during clinical trials at Johns Hopkins University Hospital has led to refinements in a commercial arrhythmia monitoring system. The performance capabilities and limitations are described for normal Coronary Care Unit use. The system evaluation spanned a time period of over 16 months and included 3,176 monitored patient days. The majority of both false and true arrhythmia alarms was found to be produced by two small, statistically independent segments of the patient population.
- Subjects :
- medicine.medical_specialty
Maryland
Remote patient monitoring
business.industry
Coronary Care Units
Biomedical Engineering
Medicine (miscellaneous)
System evaluation
Arrhythmias, Cardiac
Monitoring system
Hospital Bed Capacity, 500 and over
University hospital
Clinical trial
Electrocardiography
Patient population
Evaluation Studies as Topic
Emergency medicine
medicine
Coronary care unit
ECG analysis
business
Subjects
Details
- ISSN :
- 03638855
- Volume :
- 7
- Database :
- OpenAIRE
- Journal :
- Journal of Clinical Engineering
- Accession number :