Back to Search Start Over

Comparison of the Glidescope®, the McGrath®, the Airtraq®and the Macintosh laryngoscopes in simulated difficult airways*

Authors :
C. Abegg
Jean-Luc Waeber
Georges L. Savoldelli
François Clergue
V. Baeriswyl
Eduardo Schiffer
Source :
Anaesthesia, Vol. 63, No 12 (2008) pp. 1358-64
Publication Year :
2008
Publisher :
Wiley, 2008.

Abstract

Summary Several indirect laryngoscopes have recently been developed, but relatively few have been formally compared. In this study we evaluated the efficacy and the usability of the Macintosh, the Glidescope � , the McGrathand the Airtraqlaryngoscopes. Sixty anaesthesia providers (20 staff, 20 residents, and 20 nurses) were enrolled into this study. The volunteers intubated the trachea of a Laerdal SimManmanikin in three simulated difficult airway scenarios. In all scenarios, indirect laryngoscopes provided better laryngeal exposure than the Macintosh blade and appeared to produce less dental trauma. In the most difficult scenario (tongue oedema), the Macintosh blade was associated with a high rate of failure and prolonged intubation times whereas indirect laryngoscopes improved intubation time and rarely failed. Indirect laryngoscopes were judged easier to use than the Macintosh. Differences existed between indirect devices. The Airtraq � consistently provided the most rapid intubation. Laryngeal grade views were superior with the Airtraqand McGraththan with the Glidescope � .

Details

ISSN :
13652044 and 00032409
Volume :
63
Database :
OpenAIRE
Journal :
Anaesthesia
Accession number :