Interlaboratory comparison of SARS-CoV2 molecular detection assays in use by U.S. veterinary diagnostic laboratories

Authors :
Hon S. Ip
Sarah M. Nemser
Kirstin Frost
Shannon Pickens
Laura B. Goodman
Jodie Ulaszek
Anja Schlierf
Renate Reimschuessel
Steffen Uhlig
Robert Newkirk
Kapil Nichani
Bertrand Colson
Ravinder Reddy
Karina Hettwer
Mary Lea Killian
Andriy Tkachenko
Kaiping Deng
Matthew Kmet
Source :
Journal of Veterinary Diagnostic Investigation : Official Publication of the American Association of Veterinary Laboratory Diagnosticians, Inc
Publication Year :
2021
Publisher :
SAGE Publications, 2021.

Abstract

The continued search for intermediate hosts and potential reservoirs for SARS-CoV2 makes it clear that animal surveillance is critical in outbreak response and prevention. Real-time RT-PCR assays for SARS-CoV2 detection can easily be adapted to different host species. U.S. veterinary diagnostic laboratories have used the CDC assays or other national reference laboratory methods to test animal samples. However, these methods have only been evaluated using internal validation protocols. To help the laboratories evaluate their SARS-CoV2 test methods, an interlaboratory comparison (ILC) was performed in collaboration with multiple organizations. Forty-four sets of 19 blind-coded RNA samples in Tris-EDTA (TE) buffer or PrimeStore transport medium were shipped to 42 laboratories. Results were analyzed according to the principles of the International Organization for Standardization (ISO) 16140-2:2016 standard. Qualitative assessment of PrimeStore samples revealed that, in approximately two-thirds of the laboratories, the limit of detection with a probability of 0.95 (LOD95) for detecting the RNA was ≤20 copies per PCR reaction, close to the theoretical LOD of 3 copies per reaction. This level of sensitivity is not expected in clinical samples because of additional factors, such as sample collection, transport, and extraction of RNA from the clinical matrix. Quantitative assessment of Ct values indicated that reproducibility standard deviations for testing the RNA with assays reported as N1 were slightly lower than those for N2, and they were higher for the RNA in PrimeStore medium than those in TE buffer. Analyst experience and the use of either a singleplex or multiplex PCR also affected the quantitative ILC test results.

Details

ISSN :
19434936 and 10406387
Volume :
33
Database :
OpenAIRE
Journal :
Journal of Veterinary Diagnostic Investigation
Accession number :
edsair.doi.dedup.....db478b6e4392e52b9e01f696ca97c565
Full Text :
https://doi.org/10.1177/10406387211029913