Back to Search
Start Over
Oblique injection depth correction by a two parallel OCT sensor guided handheld SMART injector
- Source :
- Biomed Opt Express
- Publication Year :
- 2021
- Publisher :
- Optical Society of America, 2021.
-
Abstract
- We present a SMART injector with two parallel common-path optical coherence tomography fibers to enable angle measurements and injection depth corrections for oblique subretinal injection. The two optical fibers are attached to opposite sides of a 33 G needle with known offsets and designed to pass through a 23 G trocar that has an inner diameter of 0.65 mm. By attaching a SMART system to a rotational stage, the measured angles are calibrated for minimal error from reference angles. A commercial eye model was used to evaluate the control performance, and injection experiments were performed on a phantom made of agarose gel and a porcine eye.
- Subjects :
- 0303 health sciences
Optical fiber
Materials science
medicine.diagnostic_test
business.industry
Oblique case
Injector
01 natural sciences
Atomic and Molecular Physics, and Optics
Imaging phantom
Article
law.invention
010309 optics
03 medical and health sciences
Optics
Optical coherence tomography
law
0103 physical sciences
medicine
Inner diameter
business
Mobile device
030304 developmental biology
Biotechnology
Subjects
Details
- Language :
- English
- Database :
- OpenAIRE
- Journal :
- Biomed Opt Express
- Accession number :
- edsair.doi.dedup.....dd26ef0c01cf911f3f89f1546c703c3f