Back to Search
Start Over
Outcome With the Repositionable and Retrievable Boston Scientific Lotus Valve Compared With the Balloon-Expandable Edwards Sapien 3 Valve in Patients Undergoing Transfemoral Aortic Valve Replacement
- Source :
- Circulation. Cardiovascular interventions. 10(6)
- Publication Year :
- 2016
-
Abstract
- Background— New generation devices for transfemoral aortic valve replacement were optimized on valve positioning and reduction of residual aortic regurgitation. We compared 30-day, 12-month, and 24-month outcomes of the Boston Scientific Lotus valve (Lotus) and the balloon-expandable Edwards Sapien 3 (ES3) valve. Primary end point was all-cause mortality or disabling stroke within 12 months. Methods and Results— Between 2014 and 2016, 537 patients were enrolled at our center, and 202 patients received Lotus and 335 ES3. There was no residual moderate or severe aortic regurgitation. Rate of mild aortic regurgitation was lower with the repositionable and retrievable Lotus valve compared with the ES3. Rate of pacemaker implantation was significantly higher with the Lotus valve compared with the ES3 valve (36.1% versus 14.9%, P P =0.87), rate of disabling stroke (Lotus, 1.5%; ES3, 2.1%; P =0.62), or major vascular complications (Lotus, 2.9%; ES3, 2.4%; P =0.69). The primary end point at 12 months was similar between groups. In a propensity score–matched comparison, there was no difference in the primary end point within 12 months (Lotus, 15.5%; ES3, 18.6%; P =0.69) and 24 months (Lotus, 21.9%; ES3, 26.4%; P =0.49). Conclusions— Transfemoral aortic valve replacement with the ES3 and the Lotus were associated with similar 30-day, 12-month, and 24-month clinical outcomes. Need for permanent pacemaker implantation was significantly higher with the repositionable Lotus device. Clinical Trial Registration— URL: http://www.clinicaltrials.gov . Unique identifier: NCT02162069.
- Subjects :
- Aortic valve
Balloon Valvuloplasty
Male
medicine.medical_specialty
Time Factors
Lotus
Regurgitation (circulation)
Kaplan-Meier Estimate
Punctures
030204 cardiovascular system & hematology
Prosthesis Design
Transcatheter Aortic Valve Replacement
03 medical and health sciences
Disability Evaluation
0302 clinical medicine
Aortic valve replacement
Risk Factors
Internal medicine
Germany
medicine
Humans
In patient
030212 general & internal medicine
Prospective Studies
Registries
Stroke
Aged
Aged, 80 and over
biology
business.industry
Cardiac Pacing, Artificial
Hemodynamics
Arrhythmias, Cardiac
Aortic Valve Stenosis
medicine.disease
biology.organism_classification
Surgery
Femoral Artery
Stenosis
medicine.anatomical_structure
Treatment Outcome
Aortic Valve
Heart Valve Prosthesis
Cardiology
Female
Cardiology and Cardiovascular Medicine
business
Edwards sapien
Subjects
Details
- ISSN :
- 19417632
- Volume :
- 10
- Issue :
- 6
- Database :
- OpenAIRE
- Journal :
- Circulation. Cardiovascular interventions
- Accession number :