A Simple and Accurate Measurement Procedure for Serum Chloride by Ion Chromatography and Bias Evaluation of Six Ion Selective Electrode Measurements in China

Authors :
Jing Wang
Weiyan Zhou
Jiangtao Zhang
Bingqing Han
Yufei Wang
Tianjiao Zhang
Chuanbao Zhang
Jie Zeng
Ying Yan
Haijian Zhao
Source :
Clinical Laboratory. 64
Publication Year :
2018
Publisher :
Clinical Laboratory Publications, 2018.

Abstract

Background Chloride is the main anion in human body. A simple and accurate method for serum chloride measurement was developed using ion chromatography (IC) in China. Methods In the measurement, serum samples were diluted 500 times with water, filtered and injected into ion chromatography column. A mixed eluent (2 mmol/L CO32- + 12 mmol/L OH-) was used and peak area signal was collected. Five calibrators made from Standard Reference Material (SRM) 919b were used in the bracketing method. The IC method was applied as the comparative method and six ion selective electrode (ISE) measurement systems were evaluated using 60 individual patient serums. Results The IC method was proven to be accurate. The precision was 0.18% - 0.30%, the recovery was 99.66% - 100.60%, the bias was 0.19% to -0.06%, and the related expanded uncertainty was 0.775% (k = 2). The precisions of the ISE systems were smaller than the 0.9% tolerable CV except for the Beckman DXC (0.91% - 1.16%). In comparison, the results of linear regression analysis showed that the correlation coefficients were 0.9876 to 0.9979. For all systems, the range of mean biases was -5.96 - 1.48 mmol/L (-5.57% - 1.36%); the expected biases at the medical decision levels were -4.97% - 0.84% at 90 mmol/L and -6.02% - 1.76% at 120 mmol/L. All biases of the Beckman AU met the requirement of within ± 1.5%. Conclusions The IC measurement method is proven to be of high precision and trueness, and the quality of routine ISE measurement of serum chloride still needs significant improvement. The establishment of the IC method can improve the measurement quality and promote its standardization process in China.

Details

ISSN :
14336510
Volume :
64
Database :
OpenAIRE
Journal :
Clinical Laboratory
Accession number :
edsair.doi.dedup.....df7b073d8916f5f19ea3af2adc947d8c