Objective An aortic annuloplasty ring could be useful for aortic valve repair. This trial evaluated intermediate-term outcomes of internal geometric ring annuloplasty for repair of trileaflet and bicuspid aortic insufficiency associated with ascending aortic and/or aortic root aneurysms. Methods Under regulatory supervision, 47 patients with aortic insufficiency and ascending aortic (n = 22) and/or aortic root (n = 25) aneurysms were managed with aortic valve repair and aneurysm resection. Valve repair was performed using trileaflet (n = 40) or bicuspid (n = 7) internal geometric rings, together with leaflet reconstruction. Ascending aortic and/or remodeling root replacements were accomplished with Dacron grafts 5 to 7 mm larger than the rings. An Echo Core Lab provided independent echocardiographic assessments, and changes over time were evaluated by Friedman tests. Results Mean ± SD age was 60 ± 14 years, 57% (27/47) were male, 15% (7/47) had bicuspid valves, 87% (41/47) had moderate-to-severe aortic insufficiency, and 13% (6/47) had mild aortic insufficiency. All patients had annular dilatation, with a mean ± SD of 26.5 ± 2.6 mm before repair, and mean ± SD ring sizes were 21.7 ± 1.7 mm. Follow-up was 42 months (mean = 27 months). No operative mortality or valve-related complications occurred. Two patients died beyond 1 year from nonvalve-related causes. One patient required valve replacement for repair failure. Survival free of complications or valve replacement was 94% at 2 years. Significant reduction in aortic insufficiency and New York Heart Association class were observed ( P < 0.0001), and valve gradients remained low. No heart block or direct ring complications occurred. Conclusions In preliminary regulatory studies, aortic ring annuloplasty seemed safe and effective during aortic aneurysm surgery. This approach could help standardize aortic valve repair.

Details

ISSN :
15590879
Volume :
13
Issue :
4
Database :
OpenAIRE
Journal :
Innovations (Philadelphia, Pa.)
Accession number :
edsair.doi.dedup.....e251b912d7ad64c7624b282f4ef2a876