Frequency dependence of the acoustic field generated from a spherical cavity transducer with open ends

Authors :
Zhibiao Wang
Deping Zeng
Junru Wu
Faqi Li
Zhou Lin
Guangrong Lei
Dong Zhang
Min He
Song Dan
Source :
AIP Advances, Vol 5, Iss 12, Pp 127218-127218-8 (2015)
Publication Year :
2015
Publisher :
AIP Publishing LLC, 2015.

Abstract

Resolution of high intensity focused ultrasound (HIFU) focusing is limited by the wave diffraction. We have developed a spherical cavity transducer with two open ends to improve the focusing precision without sacrificing the acoustic intensity (App Phys Lett 2013; 102: 204102). This work aims to theoretically and experimentally investigate the frequency dependence of the acoustic field generated from the spherical cavity transducer with two open ends. The device emits high intensity ultrasound at the frequency ranging from 420 to 470 kHz, and the acoustic field is measured by a fiber optic probe hydrophone. The measured results shows that the spherical cavity transducer provides high acoustic intensity for HIFU treatment only in its resonant modes, and a series of resonant frequencies can be choosen. Furthermore, a finite element model is developed to discuss the frequency dependence of the acoustic field. The numerical simulations coincide well with the measured results.

Details

Language :
English
ISSN :
21583226
Volume :
5
Issue :
12
Database :
OpenAIRE
Journal :
AIP Advances
Accession number :