Back to Search
Start Over
Development and investigation of a dual-pad in-channel referencing surface plasmon resonance sensor
- Source :
- Analytical chemistry. 84(18)
- Publication Year :
- 2012
-
Abstract
- Herein, we describe the construction of a novel dual-pad referencing surface plasmon resonance (SPR) sensor utilizing electrolytic grafting of diazonium salts to individually functionalize two gold pads positioned in a single fluidic channel. Using a dove prism, a simple single axis optical train may be employed without compromising the analytical performance. Once functionalized, one pad is used as the analytical sensing pad for detection of molecular interactions while the other serves as the reference pad, compensating for background refractive index fluctuations. The reference pad effectively compensates bulk refractive index changes and temperature variations as well as other nonspecific effects. The sensor was applied to calibration of a pH-responsive polymer layer in the presence of bulk refractive index and temperature variations. Monitoring selective attachment of a protein is also demonstrated. To our knowledge, this is the first implementation of in-channel referencing SPR sensor utilizing diazonium salt-based surface chemistry.
- Subjects :
- chemistry.chemical_classification
Channel (digital image)
business.industry
Temperature
Proteins
Polymer
Equipment Design
Hydrogen-Ion Concentration
Surface Plasmon Resonance
Analytical Chemistry
Refractometry
Optics
chemistry
Dove prism
Calibration
Salts
Gold
Surface plasmon resonance
business
Layer (electronics)
Refractive index
Azo Compounds
Subjects
Details
- ISSN :
- 15206882
- Volume :
- 84
- Issue :
- 18
- Database :
- OpenAIRE
- Journal :
- Analytical chemistry
- Accession number :
- edsair.doi.dedup.....e4cbc7b3b2fad53245fc4a95c4b34593